GTM41076-06VV-Q 2MOOP (EOL due to 260-00620001(R) IC FSQ110 DIP-8 FAIRCHILD) Wall Plug-in Medical Power Supply
GTM41076-06VV-Q 2MOOP (EOL due to 260-00620001(R) IC FSQ110 DIP-8 FAIRCHILD) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
![]() | GTM41076-06VV-Q 2MOOP (EOL due to 260-00620001(R) IC FSQ110 DIP-8 FAIRCHILD), Medical Power Supply, 60601-1-4th Ed. , Wall Plug-in, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Blade Options for Q Series Wall Plug-in Power Supplies, Output Rating: 6 Watts, Power rating with convection cooling (W) , 5-30V in 0.1V increments, Approvals: cUR; PSE; PSE; EAC; Taiwan BSMI; Ukraine; CE; PSE; CB 60601-1; China RoHS; WEEE; RoHS; Double Insulation; VCCI; Class 2; NrCAN; India BIS; RCM; CCC; Level V; IP40; SIQ; CE; GTM41076-06VV-Q 2MOOP (EOL due to 260-00620001(R) IC FSQ110 DIP-8 FAIRCHILD), Medical Power Supply, 60601-1-4th Ed. , Wall Plug-in, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Blade Options for Q Series Wall Plug-in Power Supplies, Output Rating: 6 Watts, Power rating with convection cooling (W) , 5-30V in 0.1V increments, Approvals: cUR; PSE; PSE; EAC; Taiwan BSMI; Ukraine; CE; PSE; CB 60601-1; China RoHS; WEEE; RoHS; Double Insulation; VCCI; Class 2; NrCAN; India BIS; RCM; CCC; Level V; IP40; SIQ; CE; | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Information | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Model Number | GTM41076-06VV-Q 2MOOP (EOL due to 260-00620001(R) IC FSQ110 DIP-8 FAIRCHILD) | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Agency Documents | http://www.globtek.info/certs/GTM41076/ | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
CE EC-Declaration for Medical | https://cs.globtek.info/ec_declaration?id=a0Oa000000BzyHWEAZ | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
CE EC-Declaration for Industrial/Commercial/ITE | https://cs.globtek.info/ec_declaration?id=a0Oa000000BzyHWEAZ | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
CE EC-Declaration for Household | Contact GlobTek for more information | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
UKCA-Declaration for Medical | https://cs.globtek.info/ukca_declaration?id=a0Oa000000BzyHWEAZ | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
UKCA-Declaration for Industrial/Commercial/ITE | https://cs.globtek.info/ukca_declaration?id=a0Oa000000BzyHWEAZ | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
UKCA-Declaration for Household | Contact GlobTek for more information | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Morocco SDoC Declaration | https://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000BzyHWEAZ | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
RoHS/RoHS3 Declaration | https://www.globtek.com/pdf/rohs_cert/a0Oa000000BzyHWEAZ | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Prop 65 Statement | https://en.globtek.com/prop-65-statement.pdf | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
REACH Declaration | https://www.globtek.com/pdf/iso_certificates/REACH.pdf | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Conflict Minerals Declaration | https://www.globtek.com/pdf/conflict-minerals.pdf | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Model Parameters | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Type | Wall Plug-in | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Technology | Regulated Switchmode AC-DC Power Supply AC Adaptor | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Category | Medical Power Supply | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Input Voltage Range (V) | 100-240V~, 50-60 Hz | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Max Input Current (A) | 0.16A-0.1A | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Max Output Power Rating (W) | 6 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Output Voltage Range (V) | 5-30 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Efficiency | IV | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Ingress Protection | IP40 (IP52 Option Available) | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Size (mm) | 74*44*34 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Enclosure | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Rating Table | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
* (Final ordering #/Part No to be defined) | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Specifications | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
A) ELECTRICAL SPECIFICATIONS: 1. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac 2. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz 3. Output Power (Rated): 6Watts maximum 4. Output Regulation: +/- 5% measured at the output connector 5. Line Voltage Regulation: +/- 1% typical measured at the output connector 6. Output Ripple (Vp-p): +/-1% or 150 mV whichever is greater at nominal output voltage. Measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of the output connector. 7. Turn-On/Turn-Off Overshoot: 5% maximum, 500us maximum recovery time for 25% step load. 8. Turn-On Delay: 3000mSEC maximum 9. Hold-Up Time: 8mSec maximum at nominal input and full load. 10. Inrush Current: 30A typical @ 115Vac input ; 60A typical @ 230Vac input 11. Switching Frequency: 66.5 KHz typical B) PROTECTION 1. Over-Voltage: Protected unit will recover upon removal of fault 2. Short Circuit: Electronically Protected, unit will recover upon removal of fault 3. Input Protection: Input line fusing C) SAFETY 1. Dielectric Withstand Voltage: 5656Vdc from primary to secondary 2. Earth Leakage Current: N/A for Class II units, there is no PE Ground pin, so Earth Leakage current is not measured 3. Touch Current: Maximum allowed values: 100uA NC(Normal condition) 500uA SFC(single fault condition) 4. Means of Protection: 2 x MOOP D) OTHER: 1. MTBF: 200,000 hours @ 25°C ambient temperature 2. Operating Temperature: 0°C to 40°C ambient temperature 3. Humidity: 0% to 90% relative humidity 4. Storage Temperature: -40°C to 80°C E) ENCLOSURE 1. Upper Housing: High impact plastic, 94V0 polycarbonate, non-vented 2. Lower Housing: High impact plastic, 94V0 polycarbonate, non-vented 3. Size: 43.5*74.0*35.3 +/-1.0 mm | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Derating Curve | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Qty (ea.)/Carton | 80 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Individual Box/Gift Box Size (mm) | 90.5*73*53 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Individual Box/Gift Box Color | White | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Master Carton Size (mm) | 388 x 380 x 245 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Gross Weight (kg) | 15.2 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Label Width (mm) | 29.8 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Label Height (mm) | 20.8 | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Input Configuration | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Description | Blade Options for Q Series Wall Plug-in Power SuppliesData Sheet: http://en.globtek.com/interchangeable-blades.php Insertion Instructions: http://www.globtek.com/pdf/Instructions-Interchangeable-Blades.pdf Video: Q-Blade Style Instruction Video INPUT CONNECTOR: Q-Socket (below are available blades configurations which are "not included" (unless stated above); can be purchased separately, package with power supply or as a separate "Q-KIT" if specified 01. United States / Canada / Japan NEMA 1-15P/ IEC PLUG Type A [WORKS IN PLUG B] configuration: NA 2 blades, Class II; US/CA/JP P/N: Q-NA(R) Kits 01. Q-KIT: 1,2,3,4 above | ||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Output Configuration | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Common output connector options: Click to add to your request.
For a comprehensive list of options, click here Contact GlobTek for your specific requirements or custom solutions. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Approvals | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice. Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation. | |||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|