Regulated Switchmode AC-DC Power Supply AC Adaptor
Category
Medical Power Supply
Input Voltage Range (V)
100-240V~, 50-60 Hz
Max Input Current (A)
1.5A
Max Output Power Rating (W)
30
Output Voltage Range (V)
5-48
Efficiency
V
Ingress Protection
Size (mm)
101.6x50.8x22.30
Enclosure
Rating Table
Model Number (Final ordering #/Part No to be defined)
Voltage (V)
Current (A)
Power (W)
RFQ
* (Final ordering #/Part No to be defined)
Specifications
ELECTRICAL SPECIFICATIONS: 1. Input Voltage: 100-240 Vac 2. Input Current: 1.5 A RMS max 3. Input Frequency: 47 - 63 Hz 4. Output power (rated): 30Watts max Note: Refer to Power Supply information on page 2 for the corresponding rated output power for the specific output voltage. 5. Output Regulation: +/- 5% measured at the output connector 6. Line Voltage Regulation: +/- 2% typical measured at full load 7. Output Ripple (Vp-p): +/-2% maximum, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of the output connector at nominal line Optional low ripple option available (RFP4579) 8. Turn-ON/OFF Overshoot: 5% maximum, 500uS typical recovery time for 25% to 50% step load 9. Turn-ON Delay: 3 Seconds, maximum @ full load, nominal line 10. Hold-Up Time: 8mS typical @ nominal input voltage and full load 11. Inrush Current: 60A maximum cold start @ 240Vac input 12. Switching Frequency: 100 KHz typical
B) PROTECTION 1. Over-Voltage: Protected with a Zener diode clamp across the output 2. Short Circuit: Hiccup with auto recovery 3. Over-load: Hiccup with auto recovery 4. Input Protection: Input line and neutral fusing
C) SAFETY 1. Dielectric Withstand Voltage: 5656Vdc from primary to secondary 2. Warning to service personnel "CAUTION: Double pole/neutral fusing". This marking is required only for both F1 and F2 installed 3. Earth Leakage Current: <100 uA at 240Vac input voltage 4. MTBF: 200,000 hours @ 25°C ambient temperature 5. Operating Temperature: 0°C to 40°C ambient temperature
D) OTHER: 1. MTBF: 200,000 Hours @ 25°C ambient temperature 2. Humidity: 0% to 90% relative humidity 3. Storage Temperature: -10°C to 80°C 4. Cooling: Convection
Derating Curve
Qty (ea.)/Carton
Individual Box/Gift Box Size (mm)
Individual Box/Gift Box Color
Master Carton Size (mm)
TBD
Gross Weight (kg)
Label Width (mm)
Label Height (mm)
Input Configuration
Description
Solder Pins for Line & Neutral, 1.4 mm square pins, 4.5 mm long
Approvals
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.
Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.
Logo
Description
Medical 4th Ed.
EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
UKCA Certification
GMA-515496-EA
RCM certificate GMA-515958-EA; Australia and New Zealand Regulatory Compliance, Mark (http://rcm.standards.org.au/rcmfaq/rcmfaq.htm
Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [UL 62368-1:2014 Ed.2] Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [CSA C22.2#62368-1:2014 Ed.2]
LPS
Limited Power Source 62368
AAMI ES60601-1:2005 +A1 - Medical Electrical Equipment - Part 1: General Requirements CAN/CSA-C22.2 No.60601-1:14, Third Edition Issued: 2014/03/01 - Medical Electrical Equipment - Part 1 IEC 60601-1-11 Issued: 2015/01/20 Ed. 2 Medical Elec. Equip.