| USB Power Delivery Capabilities| Protocol supported: | USB Power Delivery (PD) 2.0/3.0 + PPS | 
|---|
 | Advertised Power Data Objects (PDOs/APDOs): | Standard option:     5V,, 9V, 12V, 
 
 | 
|---|
 | Note 1: | Custom fixed PDOs available upon request. PDO1 must be 5V. PDO2 through PDO7 may be set to any custom voltage from 5V to 12V | 
|---|
 | Note 2: | In critical applications, the use of a non-authorized USB PD power supply may pose a substantial risk. Power supply authorization may be implemented using USB PD Vendor Defined Messages (VDMs) to prevent system operation with non-authorized power sources. Please see our article Product Security and Risk Mitigation for USB Power Delivery (PD) Based Systems for additional information. | 
|---|
 Input| Input Voltage: | Specified: 90-264VAC, Nameplate: 100-240VAC 100% rated load current for 90-264VAC
 85% rated load current for 85-264VAC
 100% rated load current for 110-370VDC
 | 
|---|
 | Input Frequency: | Specified: 47-63Hz, Nameplate: 50-60Hz | 
|---|
 | No Load Input Power: | < 75mW @ 230VAC (EU CoC Tier 2 compliant) | 
|---|
 | Inrush Current: | < 30A @ 115VAC, < 60A @ 230VAC (cold start) | 
|---|
 | Efficiency: | DoE Efficiency Level VI and CoC Tier 2 compliant (tested according to DoE 10 CFR Part 430, Subpart B, Appendix Z) | 
|---|
 Output| Turn-on Delay: | < 1 second @ 115VAC and full load | 
|---|
 | Output Regulation | ± 4% max. (measured at the end of a 1m long output cord) | 
|---|
 | Line Regulation: | ± 0.5% typ. (measured at the end of a 1m long output cord) | 
|---|
 | Ripple: | 100mV max. (using a 47µF low-ESR electrolytic cap + 0.1µF ceramic cap, measured @ 20MHz BW, at the output connector) | 
|---|
 | Transient Response: | 5% max. deviation, 1ms max. recovery time (with 25% load step), | 
|---|
 | Hold-up Time: | 8ms min. (100VAC and full load) | 
|---|
 | Power Indicator: | Green LED | 
|---|
 Protections| Input Protection: | MOV transient suppressor, input line fusing | 
|---|
 | Over-Voltage Protection: | Level 1: 110-130%, Auto-recovery, adaptive to selected PDO Level 2: 24-28V, Latched off, cycle AC to reset
 | 
|---|
 | Over-Current Protection: | 110-130%, Auto-recovery, adaptive to selected PDO | 
|---|
 | Short-Circuit Protection: | Auto-recovery | 
|---|
 | Over-Temperature Protection: | Auto-recovery | 
|---|
 Environmental| MTBF: | 1,500,000 hours @ 25°C ambient, full load (Telcordia SR-332, Issue 3) | 
|---|
 | Operating Temperature: | -10°C to 40°C (full load) | 
|---|
 | Storage Temperature: | -40°C to 80°C | 
|---|
 | Humidity: | 0% to 95% relative humidity, non-condensing | 
|---|
 | Altitude | 5000m | 
|---|
 | Cooling: | Convection | 
|---|
 | RoHS: | Complies with EU 2011/65/EU and China SJ/T 11363-2006 | 
|---|
 Safety| Dielectric Withstand Voltage: | 4000VAC or 5656VDC from input to output 10mA, 1 minute
 | 
|---|
 | Touch Current: | NC: 80µA max. SFC: 400µA max.
 | 
|---|
 | Earth Leakage Current | 300µA max. NC/SFC (N/A for 2-conductor input models) | 
|---|
 | Means of Protection: | 2 x MOPP | 
|---|
 | Output Isolation Options: | -Q suffix: Class II 2-conductor (interchangeable blades) -T2 suffix: Class II 2-conductor
 -T3 suffix: Class II, with functional earth (FE)
 Class I, earth wire connected directly to output negative
 | 
|---|
 | Note 3: | Review output isolation options with our article:  PSU Isolation and Identify | 
|---|
 EMC| Applicable Standards: | Medical: EN 60601-1-2 (4e) Emissions: EN55032, EN61000-6-3, EN61000-6-4
 Immunity: EN55024, EN61000-6-1 (4e), EN61000-6-2 (4e)
 | 
|---|
 | Conducted Emissions: | Class B, FCC Part 15, Class B (with resistive load) | 
|---|
 | Radiated Emissions: | Class B, FCC Part 15, Class B (with resistive load) | 
|---|
 | Harmonic Current Voltage Distortion: | EN61000-3-2, Class A | 
|---|
 | Voltage Fluctuations/Flicker: | EN61000-3-3 | 
|---|
 | Electrostatic Discharge (ESD) Immunity: | EN61000-4-2, 10KV contact discharge, 18KV air discharge, Criterion A | 
|---|
 | Radiated RF Immunity: | EN61000-4-3, 10V/m @ 80-1000MHz, 3V/m @ 1-2.7GHz, 80% 1KHz AM, Criterion A | 
|---|
 | EFT/Burst Immunity: | EN61000-4-4, 4KV/100KHz., Criterion A | 
|---|
 | Line Surge Immunity: | EN61000-4-5, 2KV differential, 4KV common-mode, Criterion A | 
|---|
 | Conducted RF Immunity: | EN61000-4-6, 10VRMS, 80% 1KHz AM, Criterion A | 
|---|
 | Voltage Dip Immunity: | EN61000-4-11, Criterion B/C | 
|---|
 | Note 4: | When a Class II power supply is connected to an earth-referenced system, conducted/radiated EMI levels tend to increase since the distant earth connection creates a long return path for common-mode EMI currents. This model includes an output common-mode choke to help alleviate system-level EMI issues. This choke can be removed for cost-sensitive applications. | 
|---|
 Enclosure| Housing: | High impact plastic, 94V0 polycarbonate, non-vented | 
|---|
 | Markings: | Adhesive backed label or laser engraving | 
|---|
 Security| USB Power Delivery: | Two non-standard voltage profiles are included: 5.8V and 15.1V. System designers may use these non-standard profiles to reduce the likelihood of system operation with non-GlobTek power supplies. 12.0V is also not standard, but is often included in other power supplies. 
 The power supply will respond to a USB PD “Discover Identity” VDM with 0x4754 in the “ProductID” field. USB PD host systems may check this value before initiating/allowing power negotiation.
 | 
|---|
 | Note 5: | These measures do not guarantee a secure implementation, and are only suggested as a method of risk mitigation. | 
|---|
 | Note 6: | Please see our article Product Security and Risk Mitigation for USB Power Delivery (PD) Based Systems for additional information. | 
|---|
 Special OptionsNon-standard - Contact GlobTek Custom housing and output cord colorsCustom fixed output cord length, for applicable models (1m, 2m, 3m lengths)Custom markings and marking methodsCustom USB PD PDOs: Output voltages selectable between 5V and 20V, in 100mV increments
 |