Model Number (Final ordering #/Part No to be defined)
Voltage (V)
Current (A)
Power (W)
RFQ
GTM46101-0506-0.5-USB*
5.5
0.9
4.95
GTM46101-1306-0.7-USB*
5.2
2.45
12.99
GTM46101-1306-0.6-USB*
5.4A@2.4V
2.4
12.96
GTM46101-0506-0.9-USB*
5.1
1
5.1
GTM46101-0605-USB*
5
1.2
6
GTM46101-1005-USB*
5
2
10
GTM46101-0505-USB*
5
1
5
GTM46101-1306-0.8-USB*
5.2
2.5
13
GTM46101-1306-0.9-USB*
5.1
2.54
12.95
GTM46101-0805-USB*
5
1.5
7.5
* (Final ordering #/Part No to be defined)
Specifications
A) ELECTRICAL SPECIFICATIONS: 1. Input Voltage: Specified 90-264 Vac, Nameplate rated 100-240 Vac 2. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60 Hz 3. Output Regulation: +/- 5% measured at the end of a 1.2 meter USB cordset employing #20 Gauge conductors (RFP RFI-50856) The longest cable GlobTek recommends to be used with this product is USB cord GlobTek PN USBA1M2MICROB20(R), 1.2 Meters for best voltage regulation performance. (RFP RFI-20435-3) 4. Line Voltage Regulation: +/- 3% typical measured at the end of a 1.2 meter USB cordset employing #20 Gauge conductors 5. Output Ripple (Vp-p): 150mV Ripples shall be measured with a 47uF electrolytic capacitor paralleled with a high-frequency 0.1uF ceramic capacitor across the output. The oscilloscope's bandwidth shall be set to 20MHz. 6. Turn-On/Turn-Off Overshoot: 5% maximum, 500 us maximum recovery time for 25% step load. 7. Turn-On Delay: 3000mS MAX at 115VAC Full Load 8. Hold-Up Time: 8mS Min at 100VAC Full Load 9. Inrush Current: No damage. On power when switching on/off repeatedly under any condition. 10. Switching Frequency: 66.5 KHz typical 11. Average Efficiency: 78.7% Min.(CEC&DOE Level VI Compliant), 79.0% Min. (ErP European Commission's Ecodesign Directive (2005/32/EC) / CoC Tier 2 Compliant) 12. Input Power(no load): < 0.1W < 0.1W, < 0.075W (ErP European Commission's Ecodesign Directive (2005/32/EC) / CoC Tier 2 Compliant) 13. Cable voltage drop compensation circuitry
B) PROTECTION 1. Over-Voltage: Output shut down and auto restart 110% to 140% of Vnom 2. Over Temperature Protection: Output shut down and auto restart 3. Short Circuit Protection Output shut down and auto restart 4 Over Load Protection Output shut down and auto restart 110% to 150% of Inom 5. Input Protection: Input line fusing 6. ESD: 8 kV Contact/20kV Air Discharge
C) SAFETY 1. Dielectric Withstand Voltage: 4000Vac or 5656Vdc 10mA 1 minute 2. Earth Leakage Current: N/A for Class II units, there is no PE Ground pin, so earth Leakage current is not measured 3. Touch Current: Maximum allowed values: 100uA NC(Normal condition) 500uA SFC(single fault condition) 4. Means of Protection: 2 x MOPP 5. ROHS 3: Complies with EU 2011/65/EU amended by EU 2015/863 and China SJ/T 11363-2006
D) EMC 1. Emissions, per EN 55032, EN 61000-6-3, EN 61000-6-4 Conducted Emissions: Class B, FCC Part 15, Class B Radiated Emissions: Class B, FCC Part 15, Class B 2. Line Frequency Harmonics EN61000-3-2, Class A 3. Voltage Fluctuations/Flicker EN61000-3-3 4. Immunity, per EN 55024, EN 61000-6-1, EN 61000-6-2 Static Discharge Immunity EN61000-4-2, 8kV contact discharge, 15kV air discharge Radiated RF Immunity EN61000-4-3, 3V/m 80-2700MHz, 80% 1KHz AM. EFT/Burst Immunity EN61000-4-4, 2KV/100kHz. Line Surge Immunity EN61000-4-5, 1KV differential-mode Conducted RF Immunity EN61000-4-6, 3Vrms, 80% 1KHz AM Power Frequency Magnetic Field Immunity EN61000-4-8, NA Voltage Dip Immunity EN61000-4-11, Criteria B
E) OTHER: 1.USB2.0 , USB BC1.2 (DCP) compliant (Certain part numbers only. Check specific datasheet for details) 2. MTBF: 200,000 hours @ 25°C ambient temperature 3. Operating Temperature: 0°C to 40°C ambient temperature 4. Humidity: 0% to 95% relative humidity 5. Storage Temperature: -40°C to 80°C
F) ENCLOSURE 1. Housing: High impact plastic, 94V0 polycarbonate, non-vented 2. Markings: Label and/or Pad Printed and/or Molded in the case
INPUT CONNECTOR: Q-Socket (below are available blades configurations which are "not included" (unless stated above); can be purchased separately, package with power supply or as a separate "Q-KIT" if specified
01. United States / Canada / Japan NEMA 1-15P/ IEC PLUG Type A [WORKS IN PLUG B] configuration: NA 2 blades, Class II; US/CA/JP P/N: Q-NA(R) 02. Australian AS 3112 configuration: SAA 2 blade/ IEC TYPE I, Class II; AU P/N: Q-SAA(R) 03. UK BS 1363 configuration: UK 3 blade with dummy Ground/ IEC TYPE G, Class II; GB P/N: Q-UK(R) 04. European CEE 7/16 configuration: Europlug 2 pins/ IEC TYPE C [WORKS IN TYPE E&F], Class II; EU P/N: Q-EU(R) 05. Korean KS C8305 configuration: 2 pins/SIMILAR TO IEC TYPE C, Class II; KR P/N: Q-KR(R) 06. Argentina IRAM 2073 configuration: 2 blades/SIMILAR TO IEC TYPE I; AR P/N: Class II Q-AR(R) 07. China GB 2099 configuration: 2 blades/SIMILAR TO TYPE A, Class II; CN P/N:Q-CN(R) 08. India IS 1293 6A/BS546 configuration: 5A, 3 pins with Dummy Ground, Class II/IEC TYPE D; IN P/N: Q-IN(R) 09. IEC60320/C18 Inlet, Class II; P/N:Q-C18(R) 10. Brazilian NBR6147 configuration: 2 pins, Class II;SIMILAR TO IEC TYPE C BR P/N: Q-BR(R) 11. South Africa SABS164-1 and India IS 1293, IS694, 16A type plug, 3 round prongs, Class II + dummy ground, IEC TYPE M P/N: Q-SANS164-1-16A(R) 12. South Africa SABS164-4, 3 round prongs with a notched prong @ 0º, Class II + dummy ground, IEC TYPE M Red, P/N: Q-SANS164-4C-16A(R) 13. South Africa SABS164-4, 3 round prongs with a notched prong @ -53º, Class II + dummy ground, Blue, IEC TYPE M, P/N: Q-SANS164-4L-16A(R) 14. South Africa SABS164-4, 3 round prongs with a notched prong @ +53º, Class II + dummy ground, Black,IEC TYPE M P/N: Q-SANS164-4R-16A(R)
Contact GlobTek for your specific requirements or custom solutions.
Approvals
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.
Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.
Logo
Description
Medical 4th Ed.
EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
No Logo
SOUTH AFRICAN EMI/EMC SABS for GTM46101-1005-USB
Declaration ДС № ЕАЭС N RU Д-US.КА01.В.10453_19 Custom Union of Russia, Belarus and Kazakhstan http://www.globtek.com/redirect/?loc=gost-certificate-eac-declaration
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
No Logo
ST-Malaysia GTM46101
PRECAUCION: PARA USO EN EQUIPOS ELECTRONICOS SOLAMENTE
IRAM certificate DC-E-G71-006.1 to IEC 62368-1:2014
Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [UL 62368-1:2014 Ed.2] Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [CSA C22.2#62368-1:2014 Ed.2]
Compliance of this PSU with Industry Canada, Class B demonstrated with a standard output load. The ICES law stipulates that system-level testing is required to demonstrate compliance with the ICES-3 emission limits with the actual system load.
Compliance of this power supply with FCC Part 15, Class B has been demonstrated with a standard output load. The FCC law stipulates that system-level testing is required to demonstrate compliance with the FCC emission limits with the actual system load.
CE Certification CE Mark: tested to comply with EN 55032.2012+AC.2013 EN 61000-3-2.2014 EN 61000-3-3.2013 EN 60601-1-2.2013(Ed4) EN 55024.2010
No Logo Applicable
CB report IEC60601-1 2005 A1+C1+C2 2016-2-4 and or EN 60601-1:2006 3.1rd Edition 2xMOPP (6W max)
No Logo
CB Report IEC60950-1:2005 (2nd Edition), Am 1: 2009 + Am 2:2013
CE Mark: tested to comply with EN60601-1-2:2013 (4th Ed), EN61000-3-2, EN61000-3-3, EN50082-1, including EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6, EN61000-4-8, EN61000-4-11, EN61000-1-2-2001 EMI: EN55011 CLASS B and FCC Part 15 B
Conforms to ANSI/AAMI ES 60601-1:2005 Cert.to CAN/CSA Std. C22.2
cETLus ANSI_AAMI ES60601-1_2005
Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance [AAMI ES60601-1:2005 +A1]Medical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance [CSA C22.2#60601-1:2014
CHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
Indoor Use Only - Mark is on the label or Molded in the case
IP20
Ingress Protection: IP20 to IEC60529:2001 Protected against solids objects over 12mm and No Protection
Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
LPS
Limited Power Source 60950
No Logo Applicable
CB for IEC 62368-1:2014 (Second Edition)
RoHS
Specifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
WEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
Information Technology Equipment Safety Part 1: General Requirements (UL 60950-1 Issued: 2007/03/27 Ed: 2 Rev: 2014/10/14) Information Technology Equipment Safety Part 1: General Requirements (CSA C22.2 No. 60950-1 Issued:2007/03/27 Ed: 2 (R2012) Amd. 1
RCM certificate SAA-210063-EA Australia and New Zealand Regulatory Compliance, Mark (http://rcm.standards.org.au/rcmfaq/rcmfaq.htm)
仅适用于海拔5000米以下地区使用
CCC Altitude up to 5000 m GB17625.1-2012, GB4943.1-2011, GB/T9254-2008
Bureau Of Indian Standards for GTM46101-1005-USB (5V only)
Book 60335 up to 36 Volts
No Logo Applicable
CB to 60335-1 5th Edition (Safety of Household and Similar Electrical Appliances, Part 1: General Requirements) CB Report/ Certificate IEC 60335-1:2010 Up to 36 volts only
JAPAN TUV R-PSE, Cert. No. JD 50471624, to J62368-1(H30) , J55032(H29),J3000(H25,[15V or less]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/