GTM91099-60VV-FW Open Frame/Internal ICT / ITE / Medical Power Supply

GTM91099-60VV-FW
GTM91099-60VV-FW, ICT / ITE / Medical Power Supply,  60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Molex 09-65-2038 3 Position Header Pin 1: Neutral, Pin 2: Removed, Pin 3: Line, Output Rating: 60 Watts, Power rating with convection cooling (W) , 5-48V in 0.1V increments, Approvals: UKCA; CE; Double Insulation; RoHS; Ukraine; cUR; VCCI; WEEE; cRUus; China RoHS; CB 60601-1 2MOPP; Level V; Book 60601; CB 60950; CE; IEC/EN60950-1; South Africa; ETL; CE; S-Mark IEC/EN 60601-1 60950; EAC; IEC/EN60601-1; South Africa;GTM91099-60VV-FW, ICT / ITE / Medical Power Supply, 60601-1-4th Ed. , Open Frame/Internal, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, Molex 09-65-2038 3 Position Header Pin 1: Neutral, Pin 2: Removed, Pin 3: Line, Output Rating: 60 Watts, Power rating with convection cooling (W) , 5-48V in 0.1V increments, Approvals: UKCA; CE; Double Insulation; RoHS; Ukraine; cUR; VCCI; WEEE; cRUus; China RoHS; CB 60601-1 2MOPP; Level V; Book 60601; CB 60950; CE; IEC/EN60950-1; South Africa; ETL; CE; S-Mark IEC/EN 60601-1 60950; EAC; IEC/EN60601-1; South Africa;
Information
Model NumberGTM91099-60VV-FW
Agency Documentshttp://www.globtek.info/certs/GTM91099/
CE EC-Declaration for Medicalhttps://cs.globtek.info/ec_declaration?id=a0Oa000000C2pYVEAZ
CE EC-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ec_declaration?id=a0Oa000000C2pYVEAZ
CE EC-Declaration for HouseholdContact GlobTek for more information
UKCA-Declaration for Medicalhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000C2pYVEAZ
UKCA-Declaration for Industrial/Commercial/ITEhttps://cs.globtek.info/ukca_declaration?id=a0Oa000000C2pYVEAZ
UKCA-Declaration for HouseholdContact GlobTek for more information
Morocco SDoC Declarationhttps://cs.globtek.info/Morocco_SDoC_Declaration?id=a0Oa000000C2pYVEAZ
RoHS/RoHS3 Declarationhttps://www.globtek.com/pdf/rohs_cert/a0Oa000000C2pYVEAZ
Prop 65 Statementhttps://www.globtek.com/r2/PH7CNf
REACH Declarationhttps://www.globtek.com/r2/eM2co8
Conflict Minerals Declarationhttps://www.globtek.com/r2/vZY8R5
Model Parameters
TypeOpen Frame/Internal
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryICT / ITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)1.5 A
Max Output Power Rating (W)60
Output Voltage Range (V)5-48
EfficiencyV
Ingress Protection
Size (mm)125*75*38
EnclosureUser-added image
Rating Table
Model Number (Final ordering #/Part No to be defined)Output NumberVoltageAmps (Convection)Amps (Forced Air)Watts (Convection)Watts (Forced Air)RFQ
* (Final ordering #/Part No to be defined)
Specifications
A) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated 100-240 Vac
02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60 Hz
03. Output Regulation: +/- 5% measured at the output connector
04. Line Voltage Regulation: +/- 2% typical measured at full load
05. Output Ripple (Vp-p): +/-1% or 150mV whichever is greather at nominal output voltage. Measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with 10 uf electrolytic capacitor connected at the end of the output connector at nominal line
06. Turn-ON/OFF Overshoot: 5% maximum, 500uS typical recovery time for 25% to 50% step load
07. Turn-ON Delay: 2 Seconds, maximum @ full load, nominal line
08. Hold-Up Time: 8 mS typical @ nominal input voltage and full load
09. Inrush Current: 40/60A maximum cold start @ 115/230Vac input
10. Switching Frequency: 100 KHz typical

B) PROTECTION
1. Over-Voltage: Protected with a Zener diode clamp across the output
2. Short Circuit: Hiccup with auto recovery
3. Over-load: Hiccup with auto recovery
4. Input Protection: Input line and neutral fusing
5. Means of Patient Protection: (2 x MOPP)

C) SAFETY
01. Dielectric Withstand Voltage: 5656Vdc from primary to secondary
02. Earth Leakage Current: N/A for Class II units, there is no PE Ground pin, so Earth Leakage current is not measured
03. Touch Current: Maximum allowed values: 100uA NC(Normal condition) 500uA SFC(single fault condition)
04. ROHS 2: Complies with EU 2011/65/EU and China SJ/T 11363-2006

D) OTHER:
1. MTBF: 200,000 Hours @ 25°C ambient temperature
2. Operating Temperature: 0°C to 40°C ambient temperature
3. Humidity: 0% to 90% relative humidity
4. Storage Temperature: -10°C to 80°C
5. Cooling: Convection

E) ENCLOSURE
1. Open Frame
2. Size: 122*63.5*34 +/-1.0 mm
3. LED: Standard Green color to indicate power out (other colors available)
Derating CurveUser-added image
Qty (ea.)/Carton30
Individual Box/Gift Box Size (mm)Cell Packaging
Individual Box/Gift Box ColorN/A
Master Carton Size (mm)475 x 475 x 120
Gross Weight (kg)
Label Width (mm)80.0
Label Height (mm)7.5
Input Configuration
DescriptionMolex 09-65-2038 3 Position Header Pin 1: Neutral, Pin 2: Removed, Pin 3: LineDocument: http://www.molex.com/pdm_docs/sd/009652038_sd.pdf
Catalog Page: http://www.molex.com/molex/products/datasheet.jsp?part=active/0009652038_PCB_HEADERS.xml

User-added image
 

Mates With Part(s):

5195 SPOX™ Crimp Housing, 5239 KK® Crimp Housing
Optional Input Wiring Kits are available:

PN/Ordering No   Length  End A   End B
XHP410070M7MLX3(R)70JST XHP-4  Molex 09-93-0300
C810070M6MLX3(R)60IEC60320 C8  Molex 09-93-0300
C1410070M6MLX3(R)60IEC60320 C14  Molex 09-93-0300
Approvals

Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.

Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.

LogoDescription
Medical 4th Ed.EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
Request Info