The family is certified for IEC 60601-1 Medical Safety and EMC (4th edition, 2MOPP), IEC 60950-1 Information Technology Equipment, IEC 62368-1 Audio/video, information and communication technology equipment, UL1310 Class 2 Power Unit, and IEC 60335-1 Safety of household and similar electrical appliances. The power supplies which have an output current rating of 6A or less are all rated for Limited Power Source (LPS) application The family is certified for IEC 60601-1 Medical Safety and EMC (4th edition, 2MOPP), IEC 60950-1 Information Technology Equipment, IEC 62368-1 Audio/video, information and communication technology equipment, UL1310 Class 2 Power Unit, and IEC 60335-1 Safety of household and similar electrical appliances. The power supplies which have an output current rating of 6A or less are all rated for Limited Power Source (LPS) application
Preferred ratings shown below. Custom output voltages available in 0.1V increment from the minimum voltage listed to the maximum voltage listed; contact sales.
Model Number (Final ordering #/Part No to be defined)
Voltage (V)
Current (A)
Power (W)
RFQ
GTM96300-3015-R2*
15
2
30
GTM96300-2414.5-2.5-R2*
12
2
24
GTM96300-3014.5-2.5-R2*
12
2.5
30
GTM96300-3636-8.0-R2*
28
1.285
35.98
GTM96300-3654-R2*
54
0.666
35.96
GTM96300-107.5-2.5-R2*
5
2
10
GTM96300-2007.5-1.9-R2*
5.6
3.57
19.99
GTM96300-2419.5-4.5-R2*
15
1.6
24
GTM96300-3636-5.0-R2*
31
1.16
35.96
GTM96300-2307.5-1.8-R2*
5.7
3.94
22.46
GTM96300-2507.5-1.8-R2*
5.7
4.4
25.08
GTM96300-3648-R2_*
-48
0.75
36
GTM96300-3614.5-1.0-R2*
13.5
2.66
35.91
GTM96300-3024-4.0-R2*
20
1.5
30
GTM96300-2548-R2*
48
0.52
24.96
GTM96300-3019.5-4.5-R2*
15
2
30
GTM96300-3624-6.0-R2*
18
2
36
GTM96300-3624-5.0-R2*
19
1.894
35.99
GTM96300-3624-4.0-R2*
20
1.8
36
GTM96300-3036-6.0-R2*
30
1
30
GTM96300-3619.5-4.5-R2*
15
2.4
36
GTM96300-3619.5-1.5-R2*
18
2
36
GTM96300-2007.5-2.5-R2*
5
4
20
GTM96300-3614.5-2.5-R2*
12
3
36
GTM96300-3024-R2*
24
1.25
30
GTM96300-3048-R2*
48
0.625
30
GTM96300-2710.5-1.5-R2*
9
3
27
GTM96300-3048-11.0-R2*
37
0.811
30.01
GTM96300-3010.5-1.5-R2*
9
3.33
29.97
GTM96300-2307.5-2.5-R2*
5
4.5
22.5
GTM96300-3648-R2*
48
0.75
36
GTM96300-3648-11.0-R2*
37
0.973
36
GTM96300-3624-R2*
24
1.5
36
GTM96300-3614.5-1.5-R2*
13
2.769
36
GTM96300-2307.5-1.55-R2*
5.95
3.86
22.97
GTM96300-2307.5-R2*
7.5
3
22.5
GTM96300-3636-6.0-R2*
30
1.2
36
GTM96300-36VV-R2-XXXX
CUSTOM
CUSTOM
CUSTOM
* (Final ordering #/Part No to be defined)
Specifications
A) ELECTRICAL SPECIFICATIONS: 01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac 90-264 Vac range @ 100% of rated load current 85-264 Vac range @ 85% of rated load current 110-370 VDC range @ 100% of rated load current 02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50-60Hz 03. Output Regulation: +/- 5% measured at the output connector 04. Line Voltage Regulation: +/- 0.5% typical measured at full load 05. Output Ripple (Vp-p): 1% or 100 mV whichever is greater, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with a low impedance 47 uf electrolytic capacitor connected at the end of the output connector 06. Turn-ON/OFF Overshoot: 5% maximum, 1 mS typical recovery time for 40% to 70% step load 07. Turn-ON Delay: 1 second maximum @ full load, nominal line 08. Hold-Up Time: 8 mS typical @ nominal input voltage and full load 09. Inrush Current: 30A/60A maximum cold start @ 115/230Vac input 10. Efficiency: Compliant with Efficiency Level VI and CoC Tier 2 standards 11. No Load Standby Power: <0.075 W @ 230Vac
B) PROTECTION 01. Input Protection: Input line fusing and 300Vac MOV 02. Short Circuit/ Overload: Electronically Protected unit will auto recover upon removal of fault Output Current Limit: 110% to 160% of rated output current (upper end of current limit range is determined by normal output power rating, not derated output power level) 03. Output Over-Voltage: 110% to 130% of nominal output voltage
C) SAFETY 01. Dielectric Withstand Voltage: 4000Vac or 5656Vdc from input to output, On Class I models, 3000Vac or 4242Vdc from input to earth 02. Earth Leakage Current: Class I models < 300uA, N/A for Class II models 03. Touch Current: Class I models < 20uA, Class II models < 75uA 04. Earth Continuity Test: < 0.1 Ohm between Earth Pin at AC input and PCB termination point (Class I models only) 05. Means of Protection: 2 x MOPP 06. Compliant Standards: See listings at end of this drawing for specifics 07. Output Isolation Options: a) C8 or C18 Inlet, Class II b) C6 or C14 Inlet, Class II FE, Output Isolated from Earth contact c) C6 or C14 Inlet, Class I, Output negative directly attached to Earth contact
D) EMC EN 60601-1-2, 4th edition Emissions, per EN 55032, EN 61000-6-3, EN 61000-6-4, CISPR11 and CISPR22 Conducted Emissions: Class B, FCC Part 15, Class B Radiated Emissions: Class B, FCC Part 15, Class B Line Frequency Harmonics EN61000-3-2, Class A Voltage Fluctuations/Flicker EN61000-3-3
Immunity, per EN 55024, EN 61000-6-1, EN 61000-6-2 Static Discharge Immunity EN61000-4-2, 10kV Contact Discharge, 20kV air discharge Radiated RF Immunity EN61000-4-3, 10V/m 80-1000MHz, 3V/m 1-2.7GHz, 80% 1KHz AM. EFT/Burst Immunity EN61000-4-4, 4kV/100kHz. Line Surge Immunity EN61000-4-5, 2kV differential, 4kV common-mode Conducted RF Immunity EN61000-4-6, 3Vrms, 80% 1KHz AM Power Frequency Magnetic Field Immunity EN61000-4-8, 3A/m Voltage Dip Immunity EN61000-4-11, Performance Criteria B/C
E) OTHER: 01. MTBF: 1,000,000 Hours @ 40°C ambient temperature, Full Load 02. GTM96300 Family: -10°C to 40°C ambient temperature with full load Regarding Operating Temperature, a- See below derating table for output power capability at alternate temperature b- Extended low end temperature range available as custom option 03. Operating Humidity: 0% to 95% relative humidity, non-condensing 04. Storage Temperature: -30°C to 80°C 05. Operating Altitude: 5000 Meters 06. ROHS: Compliant with latest regulations, see approvals section below
F) ENCLOSURE 01. Housing: High impact plastic, 94V0 polycarbonate, non-vented 02. Markings: Label or Laser Printed 03. AC Input mechanical options: Desktop C6, C8, C14 or C18 IEC Inlet. Hybrid option (Desktop or Changeable Blade Wall Plug-in) Class I or Class II input
G) SPECIAL OPTIONS 01. Cost reduction, removal of LED 02. Custom Cordsets, various cordage types, and connector types 03. Custom Markings 04. Short term Output Surge Capability 05. Reduced Leakage Current versions 06. Tightened output voltage tolerance 07. Reduced output ripple level 08. Reduced output power rating marking 09. High Rel PCB laminate with Plated through Holes for IPC610 Class 2 Compliance 10. Special Housing Colors and Cordset Colors 11. Quasi Constant Current Output, for Battery Charger Applications 12. Back EMF applications, custom solutions. For unusual motor load aps and other high inductance reverse energy flow requirements 13. Improved Ingress Protection Rating
This series of Interchangeable Blade products may be used with Proprietary Interchangeable Blades as described below or with standard international power cords.
Optional INPUT BLADES: R-Socket: below are available blades configurations which are "not included" (unless stated above); may be purchased separately, packaged with power supply, or as a separate "R-KIT" if specified
Standard Cordsets Below are standard cordsets which are "not included" (unless stated above); these may be purchased separately or packaged with the power supply. Contact your Sales Engineer if the style required is not shown below. Many more available in different lengths, colors or cable material.
Contact GlobTek for your specific requirements or custom solutions.
Approvals
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.
Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.
Logo
Description
Medical 4th Ed.
EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
CCC Altitude up to 5000 m (up to 48V) GB17625.1-2012, GB4943.1-2011, GB/T9254-2008
Korea HU10499-19040A (for GTM96300-3614.5-2.5--X2 (12-14.9V)
Information Technology Equipment Safety Part 1: General Requirements (UL 60950-1 Issued: 2007/03/27Ed: 2 Rev: 2014/10/14) Information Technology Equipment Safety Part 1: General Requirements (CSA C22.2 No. 60950-1 Issued: 2007/03/27 Ed: 2 (R2012)
LPS
Limited Power Source 60950
CAN ICES-3(B)/NMB-3(B)
Compliance of this PSU with Industry Canada, Class B demonstrated with a standard output load. The ICES law stipulates that system-level testing is required to demonstrate compliance with the ICES-3 emission limits with the actual system load.
No Logo Applicable
CB for IEC 62368-1:2014 (Second Edition)
No Logo Applicable
CB for IEC 60950-1:2005 (Second Edition) + Am 1:2009 + Am 2:2013
Class 2 Power Unit
Class 2 Power Units [UL 1310:2011 Ed.6+R:01Feb2017] Power Supplies With Extra-Low Voltage Class 2 Outputs [CSA C22.2#223:2015 Ed.3] up to 48V only
IEC 60601-1-11 Issued: 2015/01/20 Ed. 2 Medical Elec. Equip.- Part 1-11: Gen. Req. for Basic Safety & Essential Perf.- Collateral Standard - Req. for Medical Elec. Equip. & Medical Elec.Systems Used in the Home Healthcare Environment; up to 48V only
No Logo Applicable
EU 230V CoC Tier 2, 278/2009, Mar 2014
Declaration ДС № ЕАЭС N RU Д-US.КА01.В.10453_19 Custom Union of Russia, Belarus and Kazakhstan http://www.globtek.com/redirect/?loc=gost-certificate-eac-declaration
Conforms to UL STD. 60950-1 Certified to CSA STD C22.2 NO.60950-1
Conforms to UL STD. 60950-1 Certified to CSA STD C22.2 NO.60950-1
Conforms to UL STD. 1310 Certified to CSA STD. C22.2 NO.223
Conforms to UL STD. 1310 Certified to CSA STD. C22.2 NO.223
Conforms to AAMI STD.ES60601-1,IEC 60601-1-11 Certified to CAN/CSA STD.C22.2 NO.60601-1
Conforms to AAMI STD.ES60601-1,IEC 60601-1-11 Certified to CAN/CSA STD.C22.2 NO.60601-1
NOM (Mexico) GTM96300-R2 NOM-029-ENER-2017 (5,7.5,12,15,19,24,48V Volts only)
IP40
Ingress Protection: ?IP40 to IEC60529:2001 Protection against granular foreign bodies
Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [UL 62368-1:2014 Ed.2] Audio/Video, Information And Communication Technology Equipment - Part 1: Safety Requirements [CSA C22.2#62368-1:2014 Ed.2]
UKCA Certification
UL1310 Warning Label Up To 42.4 Volts with plugs
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
Conforms to UL STD. 62368-1 Certified to CSA STD C22.2 NO.62368-1
RCM certificate GMA-515496-EA; Australia and New Zealand Regulatory Compliance, Mark (http://rcm.standards.org.au/rcmfaq/rcmfaq.htm
S-Mark Certificate EN 62368-1:2014 + A11 (http://www.intertek.com/marks/s/)
PRECAUCION: PARA USO EN EQUIPOS ELECTRONICOS SOLAMENTE
IRAM certificate to IEC 62368-1:2014
Compliance of this power supply with FCC Part 15, Class B has been demonstrated with a standard output load. The FCC law stipulates that system-level testing is required to demonstrate compliance with the FCC emission limits with the actual system load.
CE Certification CE Mark: tested to comply with EN 55032.2012+AC.2013 EN 61000-3-2.2014 EN 61000-3-3.2013 EN 60601-1-2.2015 EN 55024.2010
CE Certification Class II CE Mark: tested to comply with EN 55032.2012+AC.2013 EN 61000-3-2.2014 EN 61000-3-3.2013 EN 55024.2010
No Logo Applicable
CB to IEC 60335-1:2010 (Fifth Edition) incl. Corr. 1:2010 and Corr. 2:2011 + A1:2013 for up to 36V only
IEC 60601-1:2005 (Third Edition) + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) EN 60601-1:2006/ A1:2013+A12:2014 up to 48V only
CE Mark: tested to comply with EN 55032:2012+AC:2013, EN 61000-3-2:2014, EN 61000-3-3:2013, EN 55024:2010, EN 60601-1-2:2015
CHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
Indoor Use Only - Mark is on the label or Molded in the case
Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
RoHS
Specifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
Japan: Voluntary Control Council for Interference (VCCI)
WEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
JAPAN TUV R-PSE, Cert. No. JD 50473430, to J62368-1(H30) , J55032(H29),J3000(H25)[15V or less]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
JAPAN TUV R-PSE, Cert. No. JD 50473430, to J62368-1(H30) , J55032(H29),J3000(H25)[DC30? 60V]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
JAPAN TUV R-PSE, Cert. No. JD50473430 , to J62368-1(H30) , J55032(H29),J3000(H25)[DC15? 30V]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
No Logo Applicable
CB for IEC 60601-1:2005 (Third Edition) + CORR. 1 (2006) + CORR. 2 (2007) + AM1 (2012) or IEC 60601-1 (2012 reprint) (Ed 3.1) up to 48V only