Request Information Download PDF Configure Output Connector
GTM961200P120VV.V-T2 GTM961200P120VV.V-T2
Information
Model NumberGTM961200P120VV.V-T2
DescriptionGTM961200P120VV.V-T2, ITE / Medical Power Supply, 60601-1-4th Ed. , Desktop/External, Regulated Switchmode AC-DC Power Supply AC Adaptor, , Input Rating: 100-240V~, 50-60 Hz, IEC 60320/C8 AC Inlet connector, Class II, Non-Earth Ground (aka "Figure-8"), Output Rating: 120 Watts, 12-54V in 0.1V increments, Approvals: CCC ETL S-Mark CB CB CB ETL CB CB S-Mark CE China RoHS Level VI PSE Ukraine RoHS WEEE VCCI Double Insulation PSE PSE PSE
Agency Documentshttp://www.globtek.info/certs/GTM961200P/
CE EC-Declaration for Medicalhttp://globtek.force.com/models/ec_declaration?id=a0Oa000000MmWwCEAV
CE EC-Declaration for Industrial/Commercial/ITEhttp://globtek.force.com/models/ec_declaration?id=a0Oa000000MmWwCEAV
RoHS/RoHS3 Declarationhttp://www.globtek.com/pdf/rohs_cert/a0Oa000000MmWwCEAV
REACH Declarationhttp://www.globtek.com/pdf/iso_certificates/REACH.pdf
Conflict Minerals Declarationhttp://www.globtek.com/pdf/conflict-minerals.pdf
Model Parameters
TypeDesktop/External
TechnologyRegulated Switchmode AC-DC Power Supply AC Adaptor
CategoryITE / Medical Power Supply
Input Voltage Range (V)100-240V~, 50-60 Hz
Max Input Current (A)1.5 A
Max Output Power Rating (W)120
Output Voltage Range (V)12-54
EfficiencyVI
Ingress ProtectionIP41
Size (mm)149.30*62.80*34.00
EnclosureUser-added image
Rating Table
Model Number VoltageAmps (A)Watts (W) RFQ
GTM961200P11012-T2 129.2110.4 RFQ
GTM961200P12012-T2 129108 RFQ
GTM961200P11013-T2 138.54111.02 RFQ
GTM961200P11014-T2 147.93111.02 RFQ
GTM961200P12015-T2 158120 RFQ
GTM961200P12016-T2 167.5120 RFQ
GTM961200P12017-T2 177.06120.02 RFQ
GTM961200P12018-T2 186.66119.88 RFQ
GTM961200P12019-T2 196.31119.89 RFQ
GTM961200P12020-T2 206120 RFQ
GTM961200P12021-T2 215.71119.91 RFQ
GTM961200P12022-T2 225.45119.9 RFQ
GTM961200P12023-T2 235.21119.83 RFQ
GTM961200P12024-T2 245120 RFQ
GTM961200P12025-T2 254.8120 RFQ
GTM961200P12026-T2 264.61119.86 RFQ
GTM961200P12027-T2 274.44119.88 RFQ
GTM961200P12028-T2 284.28119.84 RFQ
GTM961200P12029-T2 294.13119.77 RFQ
GTM961200P12029.4-T2 29.44.081119.98 RFQ
GTM961200P12030-T2 304120 RFQ
GTM961200P12031-T2 313.87119.97 RFQ
GTM961200P12032-T2 323.75120 RFQ
GTM961200P12033-T2 333.63119.79 RFQ
GTM961200P12034-T2 343.53120.02 RFQ
GTM961200P12035-T2 353.42119.7 RFQ
GTM961200P12036-T2 363.33119.88 RFQ
GTM961200P12037-T2 373.24119.88 RFQ
GTM961200P12038-T2 383.15119.7 RFQ
GTM961200P12039-T2 393.07119.73 RFQ
GTM961200P12040-T2 403120 RFQ
GTM961200P12041-T2 412.92119.72 RFQ
GTM961200P12042-T2 422.85119.7 RFQ
GTM961200P12043-T2 432.79119.97 RFQ
GTM961200P12044-T2 442.72119.68 RFQ
GTM961200P12045-T2 452.66119.7 RFQ
GTM961200P12046-T2 462.6119.6 RFQ
GTM961200P12047-T2 472.55119.85 RFQ
GTM961200P12048-T2 482.5120 RFQ
GTM961200P12049-T2 492.44119.56 RFQ
GTM961200P12050-T2 502.4120 RFQ
GTM961200P12051-T2 512.35119.85 RFQ
GTM961200P12052-T2 522.3119.6 RFQ
GTM961200P12053-T2 532.26119.78 RFQ
GTM961200P12054-T2 542.22119.88 RFQ
SpecificationsA) ELECTRICAL SPECIFICATIONS:
01. Input Voltage: Specified 90-264 Vac, Nameplate rated: 100-240Vac
         90-264Vac range @ 100% of rated load current
         85-264 Vac range @ 90% of rated load current
         88-370 VDC range @ 100% of rated load current
02. Input Frequency: Specified 47-63 Hz, Nameplate rated 50/60Hz
03. Power Factor: 0.87 minimum @ 230Vac, 0.95 minimum @ 115Vac
04. Output Regulation: +/- 5% measured at the output connector
05. Line Voltage Regulation: +/- 0.5% typical measured at full load
06. Green Power On Indicator LED
07. Output Ripple (Vp-p): 1% or 200 mV whichever is greater, measured at 20 MHz bandwidth with 0.1 uf ceramic capacitor in parallel with a low impedance 47 uf electrolytic capacitor connected at the end of the output connector at nominal line
08. Turn-ON/OFF Overshoot: 8% maximum, 1 mS typical recovery time for 25% to 50% step load
09. Turn-ON Delay: 1 second typical
10. Hold-Up Time: 20 mS minimum
11. Inrush Current: 30A maximum at 115Vac input and 60A maximum at 230 Vac input
12. Switching Frequency: Varies from 25KHz to 125KHz
13. Efficiency: Compliant with Efficiency Level VI standard, (CoC Tier 2 option available)
14. No Load Standby Power: <0.21 W @ 230Vac, (0.15W @ 230Vac option available)

B) PROTECTION
01. Input Protection: Input line fusing and 300Vac rated MOV
02. Short Circuit/ Overload: Electronically Protected unit will auto recover upon removal of fault
         Output Current Limit: Current limited to less than 170W / Vnom and 13A
         (Vnom = nominal rated output voltage)
03. Over-Temperature : Latching Electronic Overtemperature protection. Cycle AC OFF to reset.
04. Output Over-Voltage: 110% to 135% of nominal output voltage under full load condition, and less than 60V max. Latching protection, cycle AC OFF to reset.

C) SAFETY
01. Dielectric Withstand Voltage: 4000Vac or 5656Vdc from input to output,
On Class I models, 3000Vac or 4242Vdc from input to earth
02. Earth Leakage Current: Class I models < 300uA, N/A for Class II models
03. Touch Current: Class I models < 20uA, Class II models < 70uA
04. Output Floating or Earthed option for Class I models.
05. Earth Continuity Test: < 0.1 Ohm between Earth Pin at AC input and PCB termination point (Class I models only)
06. Means of Protection: 2 x MOPP
07. Primary to Secondary Bridging Capacitance: Two IEC60384-14 certified Y1 type capacitors in series
08. Compliant Standards: See listings at end of this drawing for specifics

D) EMC
EN 60601-1-2, 4th edition
Emissions, per EN 55032, EN 61000-6-3, EN 61000-6-4
  Conducted Emissions: Class B, FCC Part 15, Class B
  Radiated Emissions: Class B, FCC Part 15, Class B
Line Frequency Harmonics EN61000-3-2, Class A
Voltage Fluctuations/Flicker EN61000-3-3
 
Immunity, per EN 55024, EN 61000-6-1, EN 61000-6-2
  Static Discharge Immunity EN61000-4-2, 10kV Contact Discharge, 18kV air discharge
  Radiated RF Immunity EN61000-4-3, 10V/m 80-1000MHz, 3V/m 1-2.7GHz, 80% 1KHz AM.
  EFT/Burst Immunity EN61000-4-4, 4kV/100kHz.
  Line Surge Immunity EN61000-4-5, 2kV differential, 4kV common-mode
  Conducted RF Immunity EN61000-4-6, 3Vrms, 80% 1KHz AM
  Power Frequency Magnetic Field Immunity EN61000-4-8, 3A/m
  Voltage Dip Immunity EN61000-4-11, Criteria

E) OTHER:
01. MTBF: 300,000 Hours @ 25°C ambient temperature
02. Operating Temperature:
    GTM961200P Family: -10°C to 40°C ambient temperature with full load
    Regarding Operating Temperature, 
    a- See below derating table for output power capability at alternate temperature
    b- Extended low end temperature range available as custom option
03. Humidity: 0% to 95% relative humidity, non-condensing
04. Storage Temperature: -30°C to 80°C
05. Cooling: Convection
06. ROHS: Compliant with latest regulations, see approvals section below
07. Operational Altitude: 5000M

F) ENCLOSURE
01. Housing: High impact plastic, 94V0 polycarbonate, non-vented
02. Size: 149.4 x 62.6 x 33.5 +/-1.0 mm
03. Markings: Label or Laser printed
04. AC Input mechanical options: Desktop C6, C8, C14 or C18 IEC Inlet.

G) SPECIAL OPTIONS
01. Custom Cordsets, various cordage types, and connector types
02. Custom Markings
03. Short term Output Surge Capability
04. Reduced Leakage Current version, medical CF leakage current compliance
05. Tightened output voltage tolerance
06. Reduced Output Ripple Level
08. Reduced output power rating
09. High Rel PCB laminate with Plated through Holes for IPC610 Class 2 Compliance
10. Special Housing Colors and Cordset Colors
Derating CurveUser-added image
Qty (ea.)/Carton20
Individual Box/Gift Box Size (mm)Egg Crate
Individual Box/Gift Box Color
Master Carton Size (mm)450 x 461 x 182
Gross Weight (kg)16
Label Width (mm)50
Label Height (mm)110
Input Configuration
DescriptionIEC 60320/C8 AC Inlet connector, Class II, Non-Earth Ground (aka "Figure-8")User-added image

Mates with IEC 60320/C7 Plug

Standard International IEC 320/C7 Cordsets

Below are standard cordsets which are "not included" (unless stated above); these may be purchased separately or packaged with the power supply. Please contact your Sales Engineer if the style required is not shown below. Many more available in different lengths, colors or cable material.

 
Conductors (No Ground) IEC 320/C7 (Figure "8")
Part NoCountryEnd AEnd BLength
4511116F703A(R)N. AmericanNEMA 1-15PIEC 320/C71830
5014112M703A(R)AustralianAS 3112IEC 320/C72000
4533501M8703(R)ChinaGB 2099.1IEC 320/C71830
2074112M703A(R)EuropeanCEE 7/16IEC 320/C72000
451J3401M8703(R)JapanJIS 8303IEC 320/C71830
2044112M703A(R)KoreaKS C 8305IEC 320/C72000
4033401M8703A(R)TaiwanCNS 10917IEC 320/C71800
6104112M703A(R)UKBS1363IEC 320/C72000
Output Configuration
Common output connector options:
Click to add to your request.






L Type (Coaxial 5.5x2.5mm plug) C Type (Coaxial 5.5x2.1mm plug) K Type (Coaxial 3.5x1.3mm plug) LL Type (5.5x2.5mm Locking 760k type) CL Type (5.5x2.1mm Locking S761k type) Magnetic Connector





YL3 Type (KPPX-3P) YL4 Type (KPPX-4P) EJ1/2/3/4/5 (EIAJ RC-5320A type connectors) MSB Type (Micro USB) Inquire for custom design

For a comprehensive list of options, click here

Contact GlobTek for your specific requirements or custom solutions.

Approvals
LogoDescription

No Logo Applicable

IEC 60950-1:2005 (Second Edition) + Am 1:2009 + Am 2:2013

No Logo Applicable

IEC 60601-1:2006/A1:2013+A12:2014 IEC 60601-1-11:2015

No Logo Applicable

IEC 60335-1:2010 (Fifth Edition) incl. Corr. 1:2010 and Corr. 2:2011 + A1:2013

No Logo Applicable

IEC 60335-1.2010+A1

No Logo Applicable

IEC 62368-1:2014 (Second Edition)
User-added imageCCC model meets both Tropical and Altitude to 5000M GB4943. 1-2011; GB/T9254-2008; GB17625. 1-2012
User-added image  CE Mark: tested to comply with EN55022:2006/A1:2007 Class B, EN610003-2, EN610003-3 including EN61000-4-2, EN61000-4-3, EN61000-4-4, EN61000-4-5, EN61000-4-6N EN61000-4-11; EMI: Complies with EN55011 CLASS B and FCC Part 15B - On label or Molded in case
User-added imageCHINA SJ/T 11364-2014, China RoHS Chart: http://en.globtek.com/globtek-rohs.php
User-added image

Double Insulation - Mark is on the label or Molded in the case
User-added imageMedical Electrical Equipment - Part 1: General Requirements For Basic Safety And Essential Performance (R2012) [AAMI ES60601-1:2005 +C1;A2] [CSA C22.2#60601-1:2014 Ed.3] [AAMI HA60601-1-11:2015 Ed.2]
User-added imageInformation Technology Equipment Safety Part 1: General Requirements >Valid without technical revision: 01Jan2022< [UL 60950-1:2007 Ed.2 +R:14Oct2014]
User-added imageIndoor Use Only - Mark is on the label or Molded in the case
User-added imageJAPAN TUV R-PSE, Cert. No. JD 50313287, to J60950-1(H26) , J55022(H22),J3000(H25)[15V or less]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
User-added imageJAPAN TUV R-PSE, Cert. No. JD 50313287, to J60950-1(H26) , J55022(H22),J3000(H25)[DC15≤30V]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
User-added imageJAPAN TUV R-PSE, Cert. No. JD 50313287, to J60950-1(H26) , J55022(H22),J3000(H25)[DC30≤60V]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
 User-added image
 
Efficiency: complies to section 301 of Energy Independence and Security Act (EISA) complies with Energy Star tier 2 (North America), ECP tier 2 (China), MEPS tier 2 (Australia), Code of Conduct (Europe)
User-added imageJAPAN TUV Rheinland-PSE GlobTek Inc to J60950-1(H26) , J55022(H22),J3000(H25).Please follow the procedure listed in the following link for proper import to Japan: http://en.globtek.com/importing-to-japan.php.
User-added image
 
Australia and New Zealand Regulatory Compliance, Mark (http://rcm.standards.org.au/rcmfaq/rcmfaq.htm
RoHSSpecifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf
User-added image
 
S-Mark Certificate EN 60601-1:2006 + A1:2013 + A12:2014 EN 60601-1-11:2015 (http://www.intertek.com/marks/s/)
User-added image
 
S-Mark Certificate EN 60950-1:2006 + A11:2009 + A1:2010 + A12:2011+A2:2013 (http://www.intertek.com/marks/s/)
User-added imageUkraine UKRSepro (Document: www.globtek.com/html/iso_certificates/GT_Ukraine.pdf)
User-added imageJapan: Voluntary Control Council for Interference (VCCI)
 User-added imageWEEE: Complies with EU 2012/19/EU (http://ec.europa.eu/environment/waste/weee/index_en.htm) Mark is on the label or Molded in the case
Popular Products
SKUVolts OutCurrent Out (Amps)Connector
358794129Female Barrel 5.5 * 2.5 * 11mm with internal Spring Clip and Locking Notch or Equal
397536129.2Female Barrel 5.5 * 2.5 * 11mm with no internal Spring Clip and with Locking Notch or Equal
358796129T-plug system
358795129Female Barrel 5.5*2.1*11mm WITH internal Spring Clip and Locking Notch or Equal
422718158Female Barrel 5.5 * 2.5 * 11mm with internal Spring Clip and Locking Notch or Equal
503064158Singatron 2DP1505-000111 or 2DP-74P06C2901T0002 with ID Resistor
358797158T-plug system
358798186.67T-plug system
377201186.67Singatron 2DP1505-000111 or 2DP-74P06C2901T0002 with ID Resistor
230497196.315-PIN ODU MINI-SNAP PC S12A1S-P05NPL0-7500 connector or equivalent
358799196.32T-plug system
358802245T-plug system
358800245Female Barrel 5.5 * 2.5 * 11mm with internal Spring Clip and Locking Notch or Equal
358801245Female Barrel 5.5*2.1*11mm WITH internal Spring Clip and Locking Notch or Equal
526771304Female Barrel 5.5*2.1*11mm WITH internal Spring Clip and Locking Notch or Equal
358803304T-plug system
397539542.22Female Barrel 5.5 * 2.5 * 11mm with internal Spring Clip and Locking Notch or Equal
358800245Female Barrel 5.5 * 2.5 * 11mm with internal Spring Clip and Locking Notch or Equal

Request Info

Text us your inquiry