GTM86100-10VV-W2U-USBC, ICT / ITE / Medical Power Supply, Wall Plug-in, AC/DC Switching Adapter with USB Power Supply AC Adaptor, , Input Rating: 100-240V~, 50/60Hz, UK BS 1363, 2 Blade w/Dummy Ground Class II, Output Rating: 10 Watts, Power rating with convection cooling (W) , 5-5.2V in 0.1V increments, Approvals:
-20°C to 60°C / 5% to 95% relative humidity, non-condensing
Cooling:
Natural convection
Safety
Certifications:
IEC60601-1 IEC60601-1-11 IEC62368-1
Dielectric Withstand Voltage:
4000VAC or 5656VDC (primary-to-secondary) 5mA, 1 minute
Output Touch Current:
NC: 100µA max. SFC: 500µA max.
Earth Leakage Current
N/A, Class II input (No protective earth pin, therefore earth leakage is not measured)
Means of Protection:
2 x MOPP (primary-to-secondary)
EMC
Certifications:
IEC60601-1-2 EN55032 (CISPR32) EN55035 (CISPR35)
Conducted Emissions:
Class B, FCC Part 15, Class B (with resistive load)
Radiated Emissions:
Class B, FCC Part 15, Class B (with resistive load)
Harmonic Current Voltage Distortion:
EN61000-3-2, Class A
Voltage Fluctuations/Flicker:
EN61000-3-3
Electrostatic Discharge (ESD) Immunity:
EN61000-4-2, 8KV contact discharge, 15KV air discharge, Criterion A
Radiated RF Immunity:
EN61000-4-3, 10V/m (80MHz-2.7GHz), Table 9, 9-28V/m (385MHz-5.78GHz ), Criterion A
EFT/Burst Immunity:
EN61000-4-4, 2KV/100kHz., Criterion A
Line Surge Immunity:
EN61000-4-5, 2KV differential, Criterion A
Conducted RF Immunity:
EN61000-4-6, 3VRMS, 80% 1KHz AM, Criterion A
Power Frequency Magnetic Field Immunity:
EN61000-4-8, N/A
Voltage Dip Immunity:
EN61000-4-11, Criterion A/B
Enclosure
Housing:
High impact plastic, PC+ABS 94V0, non-vented
Markings:
Laser marked or adhesive-backed label
Derating Curve
Qty (ea.)/Carton
300
Individual Box/Gift Box Size (mm)
43(L)*30(W)*75
Individual Box/Gift Box Color
43(L)*30(W)*75
Master Carton Size (mm)
485 x 335 x 305
Gross Weight (kg)
12
Label Width (mm)
37.0
Label Height (mm)
19.0
Input Configuration
Description
UK BS 1363, 2 Blade w/Dummy Ground Class II
The line and neutral pins have a rectangular cross section 6.35 mm by 4.0 mm, 17.7 mm long and with centres 22.2 mm apart. The protective-earth pin is a rectangular cross section 8.0 mm by 4.0 mm, 22.3 mm long and with a centre line 22.2 mm from the line/neutral pin centre line. The dimensions were originally specified in decimal inches with asymmetric tolerances and redefined as minimum and maximum metric dimensions in BS 1363:1984. Dimensions are chosen to provide safe clearance to live parts. The distance from any part of the line and neutral pins to the periphery of the plug base must be not less than 9.5 mm. This ensures that nothing can be inserted alongside a pin when the plug is in use, and helps keep fingers away from the pins. The minimum insertion distance from the face of the socket-outlet to the first point of contact with a live part is 9.6 mm. The "Dummy" longer earth pin ensures that the earth path is connected before the live pins, and remains connected after the live pins are removed. The earth pin is too large to be inserted into the line or neutral sockets by mistake.
Output Configuration
Common output connector options: Click to add to your request.