GTD9X1002-WWWVV-TW, ICT / ITE / Medical Power Supply/Class 2/Household Power Supply, 60601-1-4th Ed. , Desktop/External, DC-DC, , Input Rating: 12-36V, Lead Wire, see Blades/Cord Included for description, Output Rating: 100 Watts, Power rating with convection cooling (W) , 12-54V in 0.1V increments, Approvals: China RoHS; Double Insulation; PSE; RoHS; Ukraine; VCCI; WEEE; UKCA; GTD9X1002-WWWVV-TW, ICT / ITE / Medical Power Supply/Class 2/Household Power Supply, 60601-1-4th Ed. , Desktop/External, DC-DC, , Input Rating: 12-36V, Lead Wire, see Blades/Cord Included for description, Output Rating: 100 Watts, Power rating with convection cooling (W) , 12-54V in 0.1V increments, Approvals: China RoHS; Double Insulation; PSE; RoHS; Ukraine; VCCI; WEEE; UKCA;
ICT / ITE / Medical Power Supply/Class 2/Household Power Supply
Input Voltage Range (V)
12-36V
Max Input Current (A)
12.0A
Max Output Power Rating (W)
100
Output Voltage Range (V)
12-54
Efficiency
Ingress Protection
IP43
Size (mm)
149.4 x 62.6 x 33.5
Enclosure
Rating Table
Preferred ratings shown below. Custom output voltages available in 0.1V increment from the minimum voltage listed to the maximum voltage listed; contact sales.
Model Number (Final ordering #/Part No to be defined)
Voltage (V)
Current (A)
Power (W)
RFQ
GTD9X1002-6015-TW*
15
4
60
GTD9X1002-10022-TW*
22
4.545
99.99
GTD9X1002-10015-TP (do not use)*
15
6.666
99.99
GTD9X1002-10018-TW*
18
5.55
99.9
GTD9X1002-10012-TW*
12
8.33
99.96
GTD9X1002-10048-TW*
48
2.08
99.84
GTD9X1002-10019-TW*
19
5.263
100
GTD9X1002-10024-TW*
24
4.16
99.84
GTD9X1002-10036-TW*
36
2.77
99.72
GTD9X1002-10054-TW*
36
1.85
99.9
GTD9X1002-10028-TW*
28
3.571
99.99
GTD9X1002-10015-TW*
15
6.666
99.99
GTD9X1002-10030-TW*
48
3.33
99.9
GTD9X1002-WWWVV-TW-XXXX
CUSTOM
CUSTOM
CUSTOM
* (Final ordering #/Part No to be defined)
Specifications
Features
2 x MOPP medical isolation
Aerospace qualified per RTCA/DO-160G, EMS-ready per IEC60601-1-12
Wide "universal" DC input range
Enhanced transient and automotive/aerospace surge handling capability
Shock & vibration in accordance with IEC60601-1-12 (EMS) & DO-160G
Input line fusing (high-breaking capacity) Reverse polarity protection (up to -36VDC)
Input Over-Voltage Protection:
"Ride-through" input over-voltage protection without interruption of output voltage for high-energy surges up to 85VDC (See 'Maximum Allowable Surge Duration' plot for details. See 'EMC' section for other conditions.) Output cycles on/off IF (50V < Vin < 85V) AND IF (surge duration is outside the SOA region in the "Maximum Allowable Surge Duration" plot), 1.5s (typ.) delay between restart attempts
Input Under-Voltage Protection:
Shut down @ 7.4V (typ.), Restore @ 9.0V (max.)
Output Over-Voltage Protection:
110 - 130%, Latching, cycle DC input voltage to reset
-10°C to 40°C (100W) Check derating curve for T > 40°C.
Storage Temperature:
-40°C to 80°C
Humidity:
0% to 95% relative humidity, non-condensing
Altitude
5000m
Cooling:
Natural convection
Shock:
End use handling: Half-sine, <3ms duration, 1.65m/s, 3 shocks per face Transportation simulation: Trapezoid, 60G acceleration, 5.75m/sec, 3 shocks per face General shock: Half-sine, 100G acceleration, 6ms duration, 3 shocks per face
Vibration:
IEC60601-1-12, Section 10.1.3, Clause B DO-160G, Section 8, Procedure 8.8, Category U, Zone 1A
Bump:
EN1789, Clause 6.4.1
Drop:
EN1789, Clause 6.4.2
Ingress Protection:
IP43
RoHS:
Complies with EU 2015/863 and China SJ/T 11363-2006
Safety
Compliant Standards:
General medical: IEC60601-1 Emergency medical services (EMS): IEC60601-1-12 Information & communications technologies (ICT): IEC62368-1
Dielectric Withstand Voltage (Hi-Pot):
4300VDC from input to output, 1 minute 4300VDC from input to case, 1 minute 500VAC from output to case, 1 minute
Contact GlobTek for your specific requirements or custom solutions.
Approvals
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.
Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.
Logo
Description
Medical 4th Ed.
EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
CHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
Indoor Use Only - Mark is on the label or Molded in the case
JAPAN TUV R-PSE, Cert. No. JD 50469658, to J62368-1(H30) , J55032(H29),J3000(H25)[15V or less]. Please reference the following website for guidelines on PSE regulations: http://en.globtek.com/importing-ite-and-medical-power-supplies-ac-adaptors-to-japan/
RoHS
Specifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf