ICT / ITE / Medical Power Supply/Class 2/Household Power Supply
Input Voltage Range (V)
12-28 VDC
Max Input Current (A)
12.0A
Max Output Power Rating (W)
100
Output Voltage Range (V)
12-54
Efficiency
Ingress Protection
IP43
Size (mm)
149.4 x 62.6 x 33.5
Enclosure
Rating Table
Preferred ratings shown below. Custom output voltages available in 0.1V increment from the minimum voltage listed to the maximum voltage listed; contact sales.
Model Number (Final ordering #/Part No to be defined)
Voltage (V)
Current (A)
Power (W)
RFQ
GTD9X1003-WWWVV-TW-XXXX
CUSTOM
CUSTOM
CUSTOM
* (Final ordering #/Part No to be defined)
Specifications
Features
2 x MOPP medical isolation
EMS-ready per IEC60601-1-12
Wide "universal" DC input range
120W short-term power capability (optional)
Shock & vibration in accordance with IEC60601-1-12 (EMS)
Input
Input Voltage Range:
12 - 28VDC (nameplate/nominal)
12.4 - 30.3VDC (rated, per IEC60601-1-12) 10.5 - 33.3VDC (85% / 110% of rated, per IEC60601-1-12)
Plus short-term capability between 9.0 - 10.5V. (Duration is limited by temperature rise.)
Input Current:
9.5A (typ.) @ 12VDC 12.0A (max.) @ 10.5VDC
No Load Input Power:
< 0.5W
Efficiency:
87% @ 12VDC, 90% @ 24VDC typ., at a "typical" 60W usage scenario
Output
Turn-on Delay:
< 2 seconds
Output Regulation:
+1.5% / -4% max. (measured at the output connector)
Line Regulation:
± 0.1 % typ. (measured at the output connector)
Ripple:
1% max. (using a 10µF low-ESR electrolytic cap + 0.1µF ceramic cap, measured @ 20MHz BW, at the output connector)
-10°C to 40°C (100W) Check derating curve for T > 40°C.
Storage Temperature:
-40°C to 80°C
Humidity:
0% to 95% relative humidity, non-condensing
Altitude
5000m
Cooling:
Natural convection
Shock:
End use handling: Half-sine, <3ms duration, 1.65m/s, 3 shocks per face Transportation simulation: Trapezoid, 60G acceleration, 5.75m/sec, 3 shocks per face General shock: Half-sine, 100G acceleration, 6ms duration, 3 shocks per face
Vibration:
IEC60601-1-12, Section 10.1.3, Clause B
Bump:
EN1789, Clause 6.4.1
Drop:
EN1789, Clause 6.4.2
Ingress Protection:
IP33
RoHS:
Complies with EU 2015/863 and China SJ/T 11363-2006
Safety
Compliant Standards:
General medical: IEC60601-1 Emergency medical services (EMS): IEC60601-1-12
Dielectric Withstand Voltage (Hi-Pot):
4300VDC from input to output, 1 minute 4300VDC from input to case, 1 minute 500VAC from output to case, 1 minute
Output Leakage Current:
NC: 8µA max. SFC: 40µA max.
Means of Protection (Primary-to-Secondary):
2 x MOPP
Protection Class:
Class II (reinforced insulation)
EMC
Compliant Standards:
Medical: IEC60601-1-2
Conducted Emissions:
CISPR25, Class 2 (measured with resistive load)
Radiated Emissions:
CISPR25, Class 2 (measured with resistive load)
Electrical Transient Conduction Along Supply Lines:
ISO7637-2 (Test conditions for 12V nom. / 24V nom. systems, respectively)   Test Pulse 1: - 112V / - 450V (500 pulses), Class A   Test Pulse 2a: + 55V / + 55V (500 pulses), Class A   Test Pulse 2b: + 10V / + 20V (10 pulses), Class A   Test Pulse 3a: - 165V / - 220V (1 hour), Class A   Test Pulse 3b: + 112V / + 220V (1 hour), Class A (The above test levels correspond to "Severity Level III")
Contact GlobTek for your specific requirements or custom solutions.
Approvals
Below follows a link to the latest regulatory certifications and marks for this specific product. National standards as well as certification schemes are constantly changing and so is the set of certification marks and symbols applicable to this product. Various standards, marks and symbols, and certificates are subject to change at any time without notice.
Additional standards or unique country certifications are available upon request. Please contact Globtek to request a quotation.
Logo
Description
Medical 4th Ed.
EN/IEC 60601-1-2, 4th Edition (Please visit EMC reports for further info)
CHINA SJ/T 11364-2014, China RoHS Chart: http://www.globtek.com/pdf/F-GT-DJD-8.4.1-006%20China%20RoHS%20Declaration%205-20-22.pdf
Indoor Use Only - Mark is on the label or Molded in the case
JAPAN TUV R-PSE, Cert. No. JD 50469658, to J62368-1(H30) , J55032(H29),J3000(H25)[15V or less]. Please reference the following website for guidelines on PSE regulations: https://www.globtek.com/r2/Szj4Vb
RoHS
Specifications of directive 2011/65/EU Annex VI (ROHS-2) with amendment 2015/863-EU (ROHS-3) http://www.ce-mark.com/Rohs%20final.pdf